MAGNETIC FLOW METERS

PART 1 -- GENERAL

1.01 THE REQUIREMENT

A. The Contractor shall furnish, test, install and place in satisfactory operation the magnetic flow meters, with all spare parts, accessories, and appurtenances as herein specified and as shown on the Drawings.

1.02 RELATED WORK SPECIFIED ELSEWHERE

A. Control and Information System Scope and General Requirements

B. Powered Instruments, General

1.03 TOOLS, SUPPLIES AND SPARE PARTS

A. Furnish one portable primary head simulator for calibration and testing of magnetic flowmeter signal converters. The calibrator shall be furnished complete with rechargeable battery pack, test leads, spare battery pack, charger, carrying case and accessories. Calibrator shall be furnished by the flowmeter manufacturer, and shall be fully matched to the instrumentation furnished.

PART 2 -- PRODUCTS

2.01 MAGNETIC FLOW METER SYSTEMS

A. Magnetic flow meter systems shall include a magnetic flow tube and a microprocessor-based "smart" transmitter that is capable of converting and transmitting a signal from the flow tube. Magnetic flow meters shall utilize the characterized field principle of electromagnetic induction, and shall produce DC signals directly proportional to the liquid flow rate.

B. Each meter shall be furnished with a stainless steel or carbon steel metering tube and carbon steel flanges with a polyurethane, ceramic, neoprene, or Teflon liner as required by the application and/or as specified herein. Liner shall have a minimum thickness of 0.125 inches. The inside diameter of the liner shall be within 0.125 inches of the inside diameter of the adjoining pipe. Liner protectors shall be provided on all flow tubes.

C. The flow tube shall be provided with flush mounted electrodes. Ultrasonic electrode cleaning shall not be acceptable.

D. Grounding rings shall be provided for all meters.

E. All materials of construction for metallic wetted parts (electrodes, grounding rings, etc.) shall be minimum 316 stainless steel, but shall be compatible with the process fluid for each meter in accordance with the recommendations of the manufacturer.

F. Flow tube shall be rated for pressures up to 1.1 times the flange rating of adjacent piping. System shall be rated for ambient temperatures of ]30 to +65C. Meter and transmitter housings shall meet NEMA 4X requirements as a minimum. When meter and transmitter are located in classified explosion hazard areas, the meter and transmitter housings shall be selected with rating to meet the requirements for use in those areas. Non-metallic transmitter housings shall not be acceptable.

G. The transmitter shall provide pulsed DC coil drive current to the flow tube and shall convert the returning signal to a linear, isolated 4]20 mA DC signal. The transmitter shall utilize "smart" electronics and shall contain automatic, continuous zero correction, signal processing routines for noise rejection, and an integral LCD readout capable of displaying flow rate and totalized flow. The transmitter shall continuously run self-diagnostic routines and report errors via English language messages.

H. The transmitter's preamplifier input impedance shall be a minimum of 109-1011 ohms which shall make the system suited for the amplification of low-level input signals and capable of operation with a material build up on the electrodes.

I. The transmitter shall provide an automatic low flow cutoff below a user configurable low flow condition (0-10%). The transmitter's outputs shall also be capable of being forced to zero by an external contact operation.

J. Each flow tube shall be factory calibrated and assigned a calibration constant or factor to be entered into the associated transmitter as part of the meter configuration parameters. Manual calibration of the flow meter shall not be required. Meter configuration parameters shall be stored in non]volatile memory in the transmitter. An output hold feature shall be provided to maintain a constant output during configuration changes.

K. The transmitter shall be capable of communicating digitally with a remote configuration device via a frequency]shift]keyed, high frequency signal superimposed on the 4]20 mA output signal. The remote configuration device shall be capable of being placed anywhere in the 4]20 mA output loop. The remote configuration device shall be as specified under Section 17700. A password-based security lockout feature shall be provided to prevent unauthorized modification of configuration parameters.

L. Accuracy shall be 0.50% of rate over the flow velocity range of 0.3 to 10.0 m/s. Repeatability shall be 0.1% of rate; minimum turndown shall be 100:1. Minimum required liquid conductivity shall not be greater than 5 uS/cm. Maximum response time shall be adjustable between 1 and 100 seconds as a minimum. Transmitter ambient temperature operating limits shall be ]10 to +50C. Power supply shall be 115 VAC, 60 Hz.

M. Flow tubes shall be 150-lb flange mounted unless otherwise noted. The cables for interconnecting the meter and transmitter shall be furnished by the manufacturer. Transmitter shall be mounted integrally on flow tube, wall, or 2-inch pipe mounted as shown in the Drawings and/or as specified.

N. Magnetic flow meter systems shall be as manufactured by Rosemount or equal.

PART 3 -- EXECUTION

3.01 REQUIREMENTS

A. Ground magnetic flow meter flow tubes and grounding rings in strict accordance with the manufacturer's recommendations.

B. Refer to the specifications for further requirements.

- END OF SECTION -

Notes to Specifier:

i) Meter should be sized for 5 - 10 ft/sec velocities in pipe when used with the following process fluids:

- Primary Sludge

- RAS

- WAS

- Thickened Sludge

This will help prevent or reduce buildup of grease and oil on the electrodes.

ii) Classified areas should be specified in the instrument schedule for each meter.